The Cobas C311 Analyzer by Roche is a trusted clinical chemistry system for mid-sized laboratories. But like any advanced analyzer, it needs regular maintenance and calibration to keep running efficiently. Keeping the system clean, calibrated, and error-free ensures accurate test results, less downtime, and longer analyzer life.
This guide covers the basics of C311 maintenance, including daily and weekly tasks, calibration steps, quality control practices, and tips to troubleshoot common issues.
Daily Maintenance Tasks
Perform these simple tasks at the start or end of each day:
✔ Clean Sample and Reagent Probes
Use the system’s automatic cleaning protocols to flush probes and avoid clogs or contamination.
✔ Check Tubing and Waste Lines
Ensure tubes are not bent or blocked. Replace if there are leaks or damage.
✔ Empty the Waste Container
Always empty the waste before it reaches the full mark to avoid system errors.
✔ Wipe Down the System
Clean the outside of the analyzer with a soft, lint-free cloth. Do not use harsh chemicals.
✔ Check Reagent Levels
Make sure enough reagent is loaded for the next batch of tests. Use only Roche-approved reagent cassettes.
Weekly Maintenance Tasks
Every week, perform the following to keep the system running smoothly:
✔ Deep Clean the Mixing Cuvette
Run a cuvette washing cycle to remove residue or buildup from routine use.
✔ Prime the Fluid System
This helps clear any air bubbles and ensures consistent fluid flow.
✔ Clean ISE Module (if available)
Check and rinse the ISE electrodes, and replace the electrolyte solution if needed.
Calibration on the C311 Analyzer
What Is Calibration?
Calibration ensures that the analyzer gives correct test results by comparing to a known standard.
When to Calibrate:
After replacing reagents or test packs
After maintenance or repairs
If quality control results are out of range
On a regular schedule (e.g., weekly)
Calibration Steps:
Load Calibration Material
Use the appropriate Roche calibrator kits based on the test type.Start Calibration Procedure
On the touchscreen, select the test(s) and initiate calibration.Wait for System to Process
The system will measure the calibrators and create a calibration curve.Review and Approve
If results fall within the expected range, approve the calibration. If not, repeat or troubleshoot.Log the Calibration
Keep calibration records for lab compliance and audit purposes.
Quality Control (QC) Procedures
Purpose of QC
QC ensures test results are accurate and reliable across all patient samples.
Steps for Running QC on the C311:
Load QC material into the sample carousel.
Select QC tests from the menu.
Review results and ensure they fall within the acceptable range.
Document results and investigate any errors or outliers.
Tip: Always use two levels of QC material (normal and abnormal) for better accuracy.
Common Troubleshooting Tips
1. Clogged Sample Probe
Symptoms: Aspiration errors or missing results
Fix: Run the probe cleaning cycle, or manually flush the probe with cleaning solution.
2. Sample Not Recognized
Symptoms: Analyzer won’t start the test
Fix: Check barcode clarity, sample volume, and proper placement in the rack.
3. Reagent Error
Symptoms: Reagent not detected or test aborted
Fix: Make sure reagents are loaded properly and haven’t expired. Check the barcode for damage.
4. ISE Error (if module is installed)
Symptoms: Incorrect sodium, potassium, or chloride readings
Fix: Replace the electrolyte solution, recalibrate the ISE, and clean the electrodes.
5. Temperature or Optical Errors
Symptoms: System warning or failed tests
Fix: Check ambient lab temperature. Restart the system and contact Roche support if the issue persists.
Tips to Maximize Uptime
Follow the built-in maintenance schedule in the C311 interface
Replace worn parts as soon as errors appear
Keep backup calibrators and QC material in stock
Use only certified consumables and reagents from Roche
Keep a logbook for maintenance, QC, and troubleshooting activities
When to Call Roche Support
Contact technical support if:
Errors persist after basic troubleshooting
Optical sensors fail to calibrate
Software freezes or fails to update
You notice leaks, mechanical failure, or unknown alarms
Roche provides remote support, field service visits, and training for lab teams using the C311 analyzer.
Summary
The C311 Analyzer delivers reliable results when it’s well-maintained and regularly calibrated. With smart automation features, it minimizes manual work—but it still depends on routine care and QC checks to perform its best.
By following this maintenance guide, your lab can reduce errors, extend the analyzer’s lifespan, and keep test results consistent and compliant every day.
No comments:
Post a Comment